Camouflage - Stent Details
| SMALL VESSEL | STANDARD VESSEL | LARGE VESSEL | |
| TYPE OF DESIGN | Slotted tube, open cell design connected
with 3 interlinks per segment |
||
| EXPANSION RANGE | 2,25 mm - 2,75 mm |
3,00 mm - 4,00 mm |
4,50 mm - 6,00 mm |
| DESIGN DETAILS | 6 zigzag elements per ring segment |
6 zigzag elements per ring segment |
9 zigzag elements per ring segment |
| MATERIAL | 316 LVM |
316 LVM |
316 LVM |
| STRUT THICKNESS | 0,0033" (85 µm) |
0,0033" (85 µm) | 0,0063" (160 µm) |
| STRUT WIDTH (MAIN SEGMENT) |
0,0033" (85 µm) | 0,0039" (100 µm) | 0,0039" (100 µm) |
| STRUT WIDTH (INTERLINK) |
0,0020" (50 µm) | 0,0027" (70 µm) | 0,0027" (70 µm) |
| SHORTENING AFTER EXPANSION |
< 2 % |
< 2 % | < 4 % |
| MECHANICAL RECOIL | < 3 % | < 3 % | < 4 % |
| METAL COVERAGE | < 15 % | < 15 % | < 16 % |
| CAMOUFLAGE® LAYER THICKNESS |
100 nm |
100 nm | 100 nm |
| LONGITUDINAL FLEXIBILITY |
excellent | excellent | excellent |
| AVERAGE CROSSING PROFILE |
0,0033" (0.83 mm) | 0,0037" (0.94 mm) | 0,0053" (1.35 mm) |
| GUIDING CATHETER COMPABILITY |
5 F |
6 F |
6 F - 7 F |

DEUTSCH